Detailed investigation of to composition of fine particulate matter present in atmosphere is
necessary to understand the sources of high PM2.5/PM10 concentration in the atmosphere.
Therefore, it is necessary to not only monitor concentration of PM2.5/PM10 mass but also
required to determine the major chemical components of PM2.5/PM10 dust.
Since aerosol are typically made up of hundreds of different substances many of which are
interacting with environment and also with one another. Investigation of aerosol chemical
speciation has become a need for air pollution source
apportionment studies for air quality management &
control.
Speciation is two step sampling process of aerosol with
filter-based sampler onto a specialized filter/pretreated
cartridge followed by retrieval of samples and its
analysis in lab using sophisticated analytical
instruments.
FEATURES & SPECIALITIES:
Speciation sampler Ecotech Model AAS 406 is
specially designed for collection of PM2.5/PM10 on to
multiple 47mm dia filters followed by subsequent
laboratory analysis of collected PM2.5/PM10 dust for
different relevant chemical compounds including
EC/OC, Trace Metals, Ions (Anions & Cations)
Sulphates and Nitrates species.
AAS406 has 4 channel configuration where sampling
flow rate of 16.67 lpm is maintained using Electronic
pneumetic Mass Flow Controller at each channel independently. 4 channel speciation sampler
offers unique advantage in ease of selecting any channels for PM10 or PM2.5. Filter of any kind
can be used as per need. Its compact nature and ease of handling without compromising
effective control on monitoring duration where all 4 samples are collected for same time exactly
and simultaneously.
Each channel is provided with filter holder housing to place 47mm diameter PTFE filter/ Quartz
filter / Nylon filter/Glass micro fiber/EPM2000 filter as per sampling protocol of specific
chemical substances of interest. Provision also exists to fit Denuder for sampling of gaseous
related species, as and when required. Flow control is taking place at 16.67lpm. Filter
temperature is also recorded in logger continuously for all 4 channels. Sampler is fitted with
new generation data logger which record internal data besides user defined input data for
monitoring period.
Logger is provided with display to show important ongoing events and last recorded
information. Logger has keypad for user programming & data retrieval using USB port and pen
drive.
Logger has RS232 connectivity at which Laptop can be connected for parameter setting and for
remote access of the data. Internet is needed at Laptop with operator to allow remote user to
excess the logger data & setting through anydesk or Team Viewer. On each channel
Microprocessor based Mass Flow with Pneumetic Flow Controller is provided.
Operation and event record and sampling performance parameters like Filter Temperature,
Pressure, Volume, of air sampled, Elapsed time besides display of Ambient Temperature,
Barometeric Pressure & Relative Humidity are monitored by in formal computer fitted in
sampler.
Machine is automatically switched off if any of the following condition is arrive during
sampling to safe guard the sample.
1. If flow rate is reduced beyond range of flow controller upto 10% for 5 minutes.
2. If difference in the ambient Temp and Filter Temperature is more than 50C for 5 minutes.
3. If filter pressure variation during sampling is more than 180 mmHg.
Coefficient of variation is computed for flow rate and recorded in logger for each channel.
Sensors for recording Temperature, R.H, & Barometric Pressure are provided and data are
directly stored in Data logger at
defined interval. Additional
connector on card are also provided
to connect ultrasonic W/S & W/D
sensor. Standard unit is supplied
without W/D, W/S sensor. Only
ultrasonic W/S & W/D Sensor can be
connected need to be oredred
separately.
Ecotech product are designed to
sustain harsh environmental condition
of extreme Indian weather and dusty
atmosphere.
A compatible laptop for data
ownloading and management can
also be supplied as per requirements (optional). This need to be ordered separately. Laptop
supplied shall be loaded with 4 channel data handling software.
|