High Volume Air Samplers remain workhorse for air quality moniotring.
Modification are done in basic HVS by incorporating a cyclone in the sampler
which fractionate ambient air suspended particles into two fractions one upto 10
micron which passes through the cyclone and deposited on fitler paper while
other bove 10 micron removed by cyclone get accumulated in the cup fitted at
bottom of cyclone.
Ecotech developed RDS based on the concept and requirement given in the BIS 5182
(Part-23) 2006. Provision has been made in the RDS for fitting of Gaeous
Pollutants Sampling Attachment.
Due to provision of fitting of this attachment AAS 217 has become an unique
system using which one can sample simultanously PM10, SO2, NO2, NH3 and O3.
Collected PM10 dust can be analyzed for Heavy metals (Ni, As and Pb) and
Benzo(a)pyrene without compromizing in single dust sample collected on 8x10’’
glass micro fibre filter paper of AAS 217 as per recommended standard methods.
Detailed litreture Gaesous Pollutants Sampling Attachment Ecotech AAS 109 can be
referred for its specifications.
|