In new air quality standard notified by Govt. of India where measurement of PM10
and PM2.5 dust is required with other pollutants to decide status of air
pollution. So far samplers developed in India do not have provision for
simultaneous sampling of PM10 and PM2.5 dust. Thus two samplers need to be
operated simultaneously to know concentration of these pollutants in ambient
air. This results measurement of these parameters expensive and difficult as
significant efforts are required in transport of two samplers and their
operation in field.
Ecotech realizes these limitations and took lead to develop impactors based
Combo sampler for simultaneous sampling of PM10 & PM2.5 dust using single
instrument.
Single stage impactor matching to European Union designs has been fitted in the
Combo Sampler for PM10 measurement. This impactor removes dust greater than 10
microns from ambient air when sucked through it at constant flow rate of 2.3
m3/hr .
For sampling of 2.5 micron dust two stage impactors based on the USEPA standard
designs (published in Federal Register 40 CFR part 50 Appendix L (updated in May
2002)) has been incorporated in AAS 271.This impactor removes dust greater than
2.5 microns from ambient air in two stages( first in Pm10 impactor and second in
WINS impactor fitted in series ) when air is passed through it at flow rate of
1.0 m3/hr
Ecotech AAS 271 is fitted with Thomas make world’s best brushless and noiseless
pump assembly of adequate capacity to generate needed sampling rates
simultaneously for PM10& PM2.5 Impactor. Constant flow rates in both air stream
are maintained through out sampling duration with the help of critical orifices.
Pump assembly is equipped with heavy duty induction motor which ensures that
sampling rate are not affected due to input voltage fluctuations.
AAS 271 is equipped with two digital flow Totalizer for recording exact volume
of air sucked through filter fitted at out lets of single and two stage
impactors.
Glass Microfibre filters of 47mm dia is used for PM10 sampling while PTFE
membrane are used for PM2.5 sampling. All components of Impactor are made up of
corrosion free metals which are fabricated with the help of CNC machines to
ensure desired precision in the critical components of AAS 271 for accurate
repeatable monitoring of Pm10 & Pm2.5 dust simultaneously.
Modular design of sampler, tubing and other fitments made this sampler simple
and most suitable for simultaneous sampling PM10 &PM2.5. AAS 271 is easy to
transport , operate can be used successfully in harsh Indian climatic conditions
Performance of Combo Sampler AAS 271 has been evaluated by operating it in
parallel of two FRM USEPA certified reference samplers ( one operated for Pm10 &
another operated for Pm2.5 ) at National Environmental Engineering Research
Institute (NEERI) Nagpur. The results given by combo sampler and FRM sampler are
comparable. Thus the Pm10 & Pm2.5 values given by Combo sampler AAS 271 are
reliable and approved by NEERI Nagpur.
|